The model that was right in the laboratory and wrong in the waiting room

A glaucoma model that was accurate in validation and much less so in general practice, and how little of the trial evidence on AI in eye care measures vision.

In this issue4 sections · about 8 minutes

Between August 2021 and June 2022, researchers from the Centre for Eye Research Australia placed a self-imaging fundus camera in two general practice clinics in Victoria. An AI model read each photograph for glaucoma. They write that general practitioners "typically lack both the equipment and training necessary to diagnose glaucoma" [1]. Patients were approached while they waited to see their doctor.

The model was first trained in 2016 on colour fundus photographs from a Chinese population. In its original validation it reached a sensitivity of 95.6 percent and a specificity of 92.0 percent. However, Jan and colleagues found that it did considerably less well in the two clinics.

  • One in three people could not be photographed. Of 414 adults aged 50 or over who consented, 277 had an image captured from at least one eye. The people who could not be imaged were older, 71.4 years against 64.8, and more often reported a cataract diagnosis, 40.9 percent against 19.1 percent.
  • Sensitivity was 65.0 percent and specificity was 94.6 percent, counted per person. Counted per image, as in the original validation, sensitivity was 71.1 percent. The reference standard was two therapeutically endorsed optometrists grading the same photographs, with an ophthalmologist adjudicating. It included no optical coherence tomography (OCT) and no visual field test.
  • Most flagged patients did not attend. Among 161 people who had seen an eye care provider in the previous twelve months with no glaucoma diagnosis, the model flagged 18. Sixteen could be contacted. Six attended. One reported a glaucoma diagnosis at follow-up three months later.

The authors wrote that "too many patients in our study did not adhere to referral" [1].

Figure 1. The same glaucoma model, in the laboratory and in two Victorian general practicesSensitivity was 95.6 percent in the original validation and 65.0 percent in the clinics. Of 161 people with recent eye care and no glaucoma diagnosis, 18 were flagged, 6 attended and 1 reported a diagnosis.
A. Performance, percent
Original validationClinic, with exact 95 percent interval from the published counts
Sensitivity
Specificity
0255075100
B. Everyone who consented, people
Consented
414
Image captured, at least one eye
277
C. Subgroup: seen by an eye care provider in the previous 12 months, no glaucoma diagnosis, people
People in this subgroup
161
Flagged as referable by the AI
18
Contactable
16
Attended the referral
6
Reported a glaucoma diagnosis at follow-up
1
Jan CL et al, npj Digital Medicine 2025. Validation figures are the model's published performance in its original validation, in a different population, so the gap is not the effect of deployment alone. Clinic figures are per person, across the 277 people with an image captured from at least one eye, against two therapeutically endorsed optometrists grading the same photographs with ophthalmologist adjudication. The clinic intervals are exact binomial intervals I calculated from the paper's cross-tabulation, 13 of 20 for sensitivity and 243 of 257 for specificity. Those counts reproduce the paper's sensitivity, specificity and predictive values. They count 5 people who were ungradable on manual grading as referable, and the methods say such participants were excluded. The paper prints 59.4 to 70.6 and 91.9 to 97.2. Panel C is a subgroup and not the next stage for everyone in panel B: the 161 people who had seen an eye care provider in the previous twelve months with no glaucoma diagnosis. The referral chain is reported for that group only, and the one diagnosis is a patient report at follow-up three months later. Bars shorter than 0.6 percent of the scale are drawn at 0.6 percent so that they stay visible. [1]
Data as a table
GroupStagePeople
Everyone who consentedConsented414
Everyone who consentedImage captured, at least one eye277
Subgroup: eye care in 12 months, no glaucoma diagnosisPeople in this subgroup161
Subgroup: eye care in 12 months, no glaucoma diagnosisFlagged as referable by the AI18
Subgroup: eye care in 12 months, no glaucoma diagnosisContactable16
Subgroup: eye care in 12 months, no glaucoma diagnosisAttended the referral6
Subgroup: eye care in 12 months, no glaucoma diagnosisReported a glaucoma diagnosis at follow-up1
MeasureOriginal validationClinicExact 95% interval, clinicInterval as printed
Sensitivity95.665.040.8 to 84.659.4 to 70.6
Specificity92.094.691.0 to 97.091.9 to 97.2

The title calls it a pragmatic trial. The authors describe the design as prospective observational, with no comparator arm and no trial registration. They declare that one senior author is chief medical officer of Eyetelligence, an Australian retinal AI company. It is still the paper I would hand a colleague first, because it reports what product literature leaves out. It counts who could not be imaged, what the sensitivity was in a real clinic, and how many flagged patients turned up.

In my own practice, the referrals I remember are the ones that came back marked "did not attend".

AI is sold to optometrists on its accuracy. It is reasonable to expect an accurate tool to improve care in a primary care practice. However, this cannot be assumed, and a tool can also be in wide use without any evidence that it helps patients. This series looks at what happens after the accuracy figure. Both points apply to the tool I sell, and I come back to that at the end.

Two intervals in the paper do not reproduce

The sensitivity of 65.0 percent is printed with a 95 percent confidence interval (CI) of 59.4 to 70.6, and the specificity with 91.9 to 97.2. The paper's own table shows 15 people with referable glaucoma on manual grading and 5 more whose photographs the graders could not grade. The published figures reproduce from that table only when those 5 are counted as referable, which makes the sensitivity 13 of 20. The methods say that such participants were excluded, so the table and the methods do not agree.

An exact interval on 13 of 20 runs from 40.8 to 84.6 percent by my calculation. On 243 of 257, the specificity interval is 91.0 to 97.0. The published interval has the width that a denominator of 277 would give. Only the authors can settle it. Until they do, and whichever count is right, the true sensitivity in those clinics could plausibly be anywhere from about 40 to 88 percent.

What none of this has shown

No trial followed vision over time. I found seven randomised trials of an AI tool in ophthalmology or optometry published between January 2023 and September 2026 [2, 3, 4, 5, 6, 7, 8]. None had a vision outcome as its primary endpoint. Three measured a process of care, three measured accuracy or agreement, and one measured what patients knew and how anxious they were. The autorefraction trial measured corrected acuity on the day as a secondary outcome, and it was non-inferior.

One more trial was published on 12 October 2026, after the search window closed. Cleland and colleagues randomised 2,378 people at diabetes clinics in 24 hospitals in northern Tanzania. One group had AI screening and was told the result face to face before leaving. The other had standard screening and was told the result by telephone a few weeks later. Among the people whom masked graders found to have referable retinopathy, 35.9 percent attended the eye clinic within 8 weeks after AI screening, against 18.2 percent [9]. Its outcome was attendance, a process of care, so it does not change the count of vision outcomes.

Figure 2. The primary endpoint of each of the seven randomised trials I found, January 2023 to September 2026Three process endpoints, three accuracy or agreement endpoints, one patient-reported endpoint. No trial had vision as its primary endpoint. One measured same-day corrected acuity as a secondary outcome.
Process of care3 trials
ACCESS164 young people. Two paediatric diabetes sites, Baltimore.
Eye examination completed within 6 months100 vs 22 percent
B-PRODUCTIVE993 patients, 105 clinic days. Retina clinics, Bangladesh, randomised by clinic day.
Completed encounters per specialist hour1.59 vs 1.14
AI-OCT for macular oedema276 patients. Multicentre, Hong Kong screening programme.
False-positive referral rate24.1 vs 69.1 percent
Accuracy or agreement3 trials
Retina4IRD300 randomised, 295 analysed. Multicentre, inherited retinal disease.
Top-5 genetic category accuracy88.5 vs 67.3 percent
EyeFM copilot668 participants. Single centre, 16 ophthalmologists, China.
Correct diagnosis, and correct referral92.2 vs 75.4 percent, and 92.2 vs 80.5 percent
Deep-learning autorefraction282 adults. Crossover, Beijing Tongren Hospital.
Agreement with subjective refraction, spherical equivalentMean difference 0.07 dioptres, r = 0.96
Patient-reported1 trial
Cataract education chatbot64 patients. Single-centre pilot, Brussels.
Knowledge, anxiety and satisfactionNo significant difference detected
Vision kept over time0 trials
No trial. The autorefraction trial measured corrected acuity on the day as a secondary outcome, and it was non-inferior to subjective refraction.
Trials that randomised patients or clinic days to an AI-containing pathway and were published in the window, found by the searches described in the text. HERMES is excluded because it randomised teleophthalmology, not the AI. Each result is the AI arm against the comparator on the trial's stated primary endpoint, except the autorefraction trial, where it is the mean difference from subjective refraction and the correlation with it. Secondary outcomes were not charted for every trial, so the empty group means that no trial had vision as its primary endpoint, not that vision was never measured. The grouping by endpoint type is mine. The chatbot pilot detected no significant difference in 64 patients, which is not evidence that the two arms were equivalent. [2, 3, 4, 5, 6, 7, 8]
Data as a table
TrialParticipantsSettingPrimary endpointResult, AI arm vs comparatorEndpoint group
ACCESS164 young peopleTwo paediatric diabetes sites, BaltimoreEye examination completed within 6 months100 vs 22 percentProcess of care
B-PRODUCTIVE993 patients, 105 clinic daysRetina clinics, Bangladesh, randomised by clinic dayCompleted encounters per specialist hour1.59 vs 1.14Process of care
AI-OCT for macular oedema276 patientsMulticentre, Hong Kong screening programmeFalse-positive referral rate24.1 vs 69.1 percentProcess of care
Retina4IRD300 randomised, 295 analysedMulticentre, inherited retinal diseaseTop-5 genetic category accuracy88.5 vs 67.3 percentAccuracy or agreement
EyeFM copilot668 participantsSingle centre, 16 ophthalmologists, ChinaCorrect diagnosis, and correct referral92.2 vs 75.4 percent, and 92.2 vs 80.5 percentAccuracy or agreement
Deep-learning autorefraction282 adultsCrossover, Beijing Tongren HospitalAgreement with subjective refraction, spherical equivalentMean difference 0.07 dioptres, r = 0.96Accuracy or agreement
Cataract education chatbot64 patientsSingle-centre pilot, BrusselsKnowledge, anxiety and satisfactionNo significant difference detectedPatient-reported

The wider AI trial literature to November 2023 looks much the same. A scoping review of 86 randomised trials of AI in clinical practice found that 70, or 81 percent, had positive primary endpoints. In 46 of the 86, or 54 percent, the endpoint was diagnostic yield or performance. Its full text records that all three ophthalmology trials reported a reduction in operational time [10].

The count comes from the evidence map behind this series. The map was built from PubMed, ClinicalTrials.gov, the Australian New Zealand Clinical Trials Registry and the regulator lists between 27 and 30 September 2026. It is not a registered systematic search, so a trial indexed under terms I did not use would be missing.

Two more gaps run through this literature. Attendance after a positive result is the step the software cannot take for the patient. It was 16 of 25 young people in a United States diabetes clinic trial [2], 6 of the 16 who could be reached in the glaucoma clinics, and just over one in three people with referable retinopathy in the AI group in Tanzania.

Outside diabetic retinopathy the evidence is also thin. The only OCT model I found tested in community optometry, in the United Kingdom, could process images from only half the participants and was no more accurate than the optometrists [11].

When the clinician and the AI disagree

Two studies measured what an AI reading does to the clinician's own decision.

Explaining a glaucoma AI's recommendation did not measurably improve optometrists' accuracy. One kind of explanation made them more likely to follow the AI when it was wrong. Gomez and colleagues put 87 optometrists in front of a glaucoma referral model built from routine eye care data. Accuracy was 51 percent unaided (95 percent CI 46 to 56). It was 58 percent with the AI recommendation and no explanation, 60 percent with a feature-importance explanation and 62 percent with a scorecard explanation. The three AI-supported conditions did not differ significantly from one another, so the study found no gain from the explanations.

Post-hoc explanations "increased over-reliance on incorrect AI recommendations" [12]. The authors note that the effect was small and rested on few cases.

The investigators chose the 20 cases so that the AI was right on 16 of them. The 80 percent describes the case set and not the model in a clinic.

In a second study, Kashiwagi and colleagues asked 45 trainee and expert ophthalmologists to grade 120 fundus photographs. At least a week later they read the same photographs again with a simulated AI. They were told it was highly accurate, and its answers were set to be wrong 30 percent of the time. Accuracy rose from 48.4 to 59.6 percent overall. It was 63.9 percent when the AI was right and 47.9 percent when it was wrong [13].

However, the paper tested the readings made with a wrong AI against the first-session readings and found no significant difference. The 47.9 percent covers a subset of the photographs and the 48.4 percent covers all 120, so those two figures are not a matched pair. The gain was larger for trainees, 14.2 percentage points against 8.6, and the effect of a wrong AI was less pronounced in experts. The standard deviations run from 20 to 27 percentage points, and a repeat reading at least a week apart carries its own learning effect.

Figure 3. Clinicians with AI advice: better on average, and no better when the advice was wrongOptometrists with AI scored below the AI on the study cases. Ophthalmologists scored 63.9 percent when the AI was right and 47.9 percent when it was wrong.
A. 87 optometrists, urgent glaucoma referral, percent correct
B. 45 ophthalmologists, glaucoma grading, all 120 photographs, percent correct
C. The second reading, split by whether the AI was right, percent correct
Panel A: a user study of 87 optometrists with a glaucoma referral model. The published 95 percent intervals are 46 to 56 unaided, 56 to 60 with no explanation, 55 to 64 with feature importance and 57 to 66 with a scorecard, and the three AI-supported conditions did not differ significantly. The AI figure is its accuracy on the 20 cases selected for the study (Gomez 2025). Panels B and C: 45 trainee and expert ophthalmologists grading 120 fundus photographs unaided, then the same photographs one week later with an AI whose judgements were deliberately wrong 30 percent of the time. Standard deviations are 24.8, 20.3, 20.6 and 26.6 points in the order shown, and a standard deviation is not a confidence interval (Kashiwagi 2025). Panel C splits the second reading into two subsets. The 47.9 percent is on the photographs where the AI was wrong and the 48.4 percent is on all 120, so the two are not a matched pair. The paper's own test of the wrong-AI readings against the first-session readings found no significant difference. [12, 13]
Data as a table
87 optometristsAccuracy, percentPublished 95% interval
Optometrists alone5146 to 56
With AI, no explanation5856 to 60
With AI, feature importance6055 to 64
With AI, scorecard6257 to 66
AI alone, on the 20 study cases80not reported
45 ophthalmologistsPhotographsAccuracy, percent
First reading, unaidedAll 12048.4
Second reading, with AIAll 12059.6
Second reading, with AIPhotographs where the AI was right63.9
Second reading, with AIPhotographs where the AI was wrong47.9

Conflict of interest, funding and how this was written

I own OptoGuide Pty Ltd, which sells OptoGuide, a clinical decision support product for optometrists. Its clinical recommendations come from fixed rules, and no AI model runs when a clinician uses it.

No one funded this issue. I have no financial relationship with Eyetelligence, the Centre for Eye Research Australia, or any maker of an AI system named here. I checked every figure against the paper or the register it is cited to, and the errors are mine. This series will cover the field without promoting OptoGuide, and where a finding bears on rule-based decision support I will say so.

Before I believed OptoGuide changed a patient's outcome, I would want to see referral decisions made with and without it, in clinics I do not run, and followed through to what happened to each patient.

Issue 3 follows the diabetic retinopathy systems out of their pivotal trials: what held, what moved, who could be photographed and who came back after a positive screen. Issue 4 maps where AI now sits in an eye examination, what each tool does and which tools are on the Australian register. The issue after that covers the rules for Australian optometrists using AI: the guidance from the Australian Health Practitioner Regulation Agency (Ahpra) and the Code of conduct, the TGA's treatment of software as a medical device, the privacy obligations, including the automated decision-making transparency requirement that starts on 10 December 2026, and where liability sits when a tool is wrong.

A scoping review of this literature is in preparation. Each issue of Pixel & Pupil is commentary on a subset of the studies. The review is a separate piece of work, and its protocol will be registered before the searches are run.

References

  1. Jan CL, Joseph S, Vingrys AJ, et al. Prospective pragmatic trial of automated retinal photography and AI glaucoma screening in Australian primary care. npj Digit Med. 2025;8(1):386. doi:10.1038/s41746-025-01768-y. PMID 40595445.
  2. Wolf RM, Channa R, Liu TYA, et al. Autonomous artificial intelligence increases screening and follow-up for diabetic retinopathy in youth: the ACCESS randomized control trial. Nat Commun. 2024;15(1):421. doi:10.1038/s41467-023-44676-z. PMID 38212308. NCT05131451.
  3. Abràmoff MD, Whitestone N, Patnaik JL, et al. Autonomous artificial intelligence increases real-world specialist clinic productivity in a cluster-randomized trial. npj Digit Med. 2023;6(1):184. doi:10.1038/s41746-023-00931-7. PMID 37794054. NCT05182580.
  4. Zhang S, Ran A, Zhou J, et al. An AI-based OCT system to detect diabetic macular edema: a prospective validation and noninferiority randomized clinical trial. JAMA. 2026;336(3):215-223. doi:10.1001/jama.2026.7025. PMID 42295755. ChiCTR2300075087.
  5. Jia H, Qian B, Qu Y, et al. AI-based clinician decision support system for diagnosis of inherited retinal diseases: a multicenter, randomized trial. Nat Med. 2026;32(9):3214-3223. doi:10.1038/s41591-026-04545-w. PMID 42498742. NCT06839170.
  6. Wu Y, Qian B, Li T, et al. An eyecare foundation model for clinical assistance: a randomized controlled trial. Nat Med. 2025;31(10):3404-3413. doi:10.1038/s41591-025-03900-7. PMID 40877476. ChiCTR2500095518.
  7. Yan H, Yi Z, Qilin S, et al. Deep learning-empowered low-cost portable automated refraction system: A solution to the inadequate effective correction rate of refractive errors in resource-limited areas. Ophthalmic Physiol Opt. 2025;45(7):1676-1688. doi:10.1111/opo.70027. PMID 41133975.
  8. Cardon de Lichtbuer P, Declerck W, Rosseel Q, et al. Enhancing patient education in cataract surgery using a conversational artificial intelligence chatbot: pilot randomized controlled trial. J Cataract Refract Surg. 2026;52(5):470-475. doi:10.1097/j.jcrs.0000000000001856. PMID 41408706.
  9. Cleland CR, Bascaran C, Makupa WU, et al. Artificial intelligence-supported diabetic retinopathy screening in Tanzania: a randomised controlled trial. Lancet Glob Health. 2026;104062. doi:10.1016/j.langlo.2026.104062. PMID 42863478. ISRCTN18317152.
  10. Han R, Acosta JN, Shakeri Z, et al. Randomised controlled trials evaluating artificial intelligence in clinical practice: a scoping review. Lancet Digit Health. 2024;6(5):e367-e373. doi:10.1016/S2589-7500(24)00047-5. PMID 38670745.
  11. Sharma A, Hussain R, Learoyd AE, et al. Tele-ophthalmology-enabled and artificial intelligence-ready referral pathway for community optometry referrals of retinal disease: HERMES cluster randomised trial with a diagnostic accuracy study. Health Technol Assess. 2025;29(69):1-113. doi:10.3310/QNDF3325. PMID 41424155. ISRCTN18106677.
  12. Gomez C, Wang R, Breininger K, et al. The explainable AI dilemma under knowledge imbalance in specialist AI for glaucoma referrals in primary care. npj Digit Med. 2025;8(1):706. doi:10.1038/s41746-025-02069-0. PMID 41266539.
  13. Kashiwagi K, Toyoura M, Mao X, et al. Influence of artificial intelligence on ophthalmologists’ judgments in glaucoma. PLoS One. 2025;20(4):e0321368. doi:10.1371/journal.pone.0321368. PMID 40238811.

Pixel & Pupil is written by Ankit Mathur and published by OptoGuide Pty Ltd. It is not advertising, carries no testimonials and accepts no sponsorship. Figures are drawn from the published values cited beneath each one.